- Collect stool sample by using the sample collection device provided.
- Unscrew the top of the sample collection device, take out the sample collection stick, and collect the sample by dipping the stick into 3~6 different places of the stool sample.
- Put the sample collection stick back in the sample collection device and screw together tightly.
- If the sample cannot be tested on the day of collection, store the stool sample at 4°C. Bring the specimen to room temperature before testing.

TEST PROCEDURE:
- Remove the test device from the sealed pouch.
- Specimen collection. Please see also “SPECIMEN COLLECTION AND STORAGE&rdquo.
- Shake the sample collection tube several times.
- Holding the sample collection tube upright, carefully unscrew the top of collection tube.
- Squeeze 2-3 drops of sample solution into the sample well “S” .
- Read results between 5 and 10 minutes. Do not read results after 10 minutes.

INTERPRETATION OF RESULTS:
Positive: Two distinct pink-colored lines appear, one in the test region (T) and one in the control region(C). This indicates a positive result no matter the test region(T) is stronger or weaker than the control region(C).
Negative: Only one colored line appears in the control region(C). This indicates a negative test result.
Invalid: Control Line fails to appear. The test result is invalid regardless of the presence or absence of the test region(T).
NOTE:Insufficient specimen volume or incorrect procedural techniques are the most likely reasons for control line failure.Review the procedure and repeat the test with a new test strip. If the problem persists, please contact your local distributor.
SPECIFICATION:
Description | Specimen | Format | Catalog No. | Sensitivity | Package Size | Strip Width | Certificate |
(TRF) Transferrin Test | Feces | Cassette | H20-20 | 40ng/mL | 15 Tests | 4.0 mm | —— |