- Bring the pouched test device to room temperature(15-30℃) prior to testing. Do not open the pouch until ready to perform the assay.
- Remove the test device from the sealed pouch. Lay it on a flat, clean and dry surface.
- Use the pipette to draw and slowly add 1 drop of whole blood to the sample well.
- Hold the buffer bottle vertically and add 1-2 drops to the sample well./ If using a pipette, change a new one to avoid cross-contamination. Draw and transfer 2-3 drops of buffer to the sample well.
- Interpret test results within 10-15 minutes. Do not interpret after 20 minutes.
INTERPRETATION OF RESULTS:

INTENDEDUSE:
IgG Positive:
The control line and IgG line (T1) are visible in the result window. The test is positive for IgG antibodies. This indicates a past dengue infection.
IgM Positive:
The control line and IgM line (T2) are visible in the result window. The test is positive for IgM antibodies. This indicates a primary dengue infection.
IgM and IgG Positive:
The control line, IgM (T2)and IgG (T1) lines are visible in the result window. The test is positive for IgM and IgG antibodies. This indicates a secondary dengue infection.
Negative:
The control line is the only line visible in the result window. No IgG or IgM antibodies have been detected. The result does not exclude dengue infection. If symptoms persist, a new specimen should be drawn from the patient in 3-5 days and then should be retested (see the limitations section).
Invalid:
If the control line does not appear in the result window, the test results are INVALID regardless of the presence or absence of the line in the test region.
NOTE: Insufficient specimen volume or incorrect procedural techniques are the most likely reasons for the control line failure. Review the procedure and repeat the test with a new device. If problem persists, please contact your local distributor.
SPECIFICATION:
Description | Specimen | Format | Catalog No. | Sensitivity | Package Size | Strip Width | Certificate |
Dengue lgG/gM Rapid test | Serum/Plasma | Cassette | B30-20 | 94.4% | 25 Tests | 4.0 mm | CE |
Whole Blood | Cassette | B30-21 | 25 Tests | 4.0 mm | |||
B30-22 | 1 Test | 4.0 mm |