Strip

 

  Product Name:           
  Malaria P.f & P.v Antibody Test (Strip) 
  Accessories:          
  Pipette, instruction for use, buffer         
  Principle of Test:           
  Immunochromatographic assay
  Product Performance:
  Sensitivity: 91.3%

 




INTENDED USE: Malaria P.f & P.v Ab test is an immunochromatographic (rapid) test for the qualitative detection of antibodies of all isotypes (IgG, IgM, IgA) specific to Plasmodium falciparum and Plasmodium vivax simultaneously in human serum or plasma.

TEST PROCEDURE

1. Bring the pouched test strip to room temperature(15-30℃) prior to testing. Do not open pouch until ready to 

perform the assay. 

2. Remove the test strip from the sealed pouch. Lay it on a flat, clean and dry surface.

3. Hold the dropper vertically and transfer 1 drop of serum or plasma to the sample pad of the test strip. Then hold 

the buffer bottle vertically and add one drop and start the timer.

4. Wait for the red line to appear. Read result between 10-20minutes. Do not interpret the result after 20 minutes.

INTERPRETATION OF RESULTS   

                                                                

P.f Positive    P.v Positive      P.f and P.v Positive        Negative        Invalid 

 

Malaria P.f Positive

The control line and Malaria P.f line (T1) are visible in the result area. The test is positive for Malaria P.f. 

Malaria P.v Positive 

The control line and Malaria P.v line (T2) are visible in the result area. The test is positive for Malaria P.v. 

Malaria P.f and Malaria P.v Positive

The control line, Malaria P.f (T1) and Malaria P.v (T2) lines are visible in the result area. The test is positive 

for Malaria P.f and Malaria P.v. 

 

Negative 

The control line is the only line visible in the result area. No Malaria P.f or Malaria P.v has been detected. 

Invalid  

If the control line does not appear in the result area, the test results are INVALID regardless of the presence 

or absence of the line in the test region. 

NOTE: Insufficient specimen volume or incorrect procedural techniques are the most likely reasons for the control 

line failure. Review the procedure and repeat the test with a new strip. If problem persists, please contact 

your local distributor.